Process calibrator: mA,V,mV,Ω,Pt100,Cu50, TC:E/K/B/S/T/R/J/N Maximize

Process calibrator: mA,V,mV,Ω,Pt100,Cu50, TC:E/K/B/S/T/R/J/N

This instrument can output/measurement the following  signals:Current/Voltage/mV; thermocouple (E,K,B,S,T,R,J,N), resistance 999.9 Ω, Pt100,Cu50, containing 14 input,15 output,analog meter site can  be read or the vast majority of quantities.V,mV,mA, Ω input and output has a  corresponding button,call directly,no combination of keys,quick and easy  operation.Small instrument,Rich functionality

More details

LB02 Process Calibrator

1 Item in stock

318,00 лв

Power supply
battery rechargeable 
(or AC power adapter not included)
Battery
Approx. 20 hours under the condition of 10mA
Max.allowable voltage
30V
Operation temperture range
0-50°C
Operation humidity range
≤80%RH
Store temperature range
≤-10°C-50°C
Store humidity range
≤90%RH
Size
120×80×25mm
Weight
350g(with holster 150x100x40mm)
Accessories
Operation manual,test line(probe)
Options
AC adapter(VCPS), not incluided
Security
Conforms to IEC 1010
 
INPUT
Messure type
Input range
Resolution
Accuracy
Remark
DC Voltage
V
 0 ~30.000V
 0.001V
0.025%+4
 
DC milivolt
mV
0~150.00mV
0.01mV
0.025%+2
 
Ohm
Ω
0~999.9 Ω
0.1 Ω
0.025%+3
Not contain lead resistance
DC Current
mA
0~30.000mA
0.001mA
0.025%+4
 
Thermocouple
R
0~1700 °C
1°C
0.1%+3
Not include cold junction compensation error
 
N
0~1300 °C
1°C
0.1%+3
Not include cold junction compensation error
 
S
0~1600 °C
1°C
0.1%+3
Not include cold junction compensation error
 
K
0~1370 °C
1°C
0.1%+1
Not include cold junction compensation error
 
E
0~1000 °C
1°C
0.1%+1
Not include cold junction compensation error
 
J
0~1200 °C
1°C
0.1%+1
Not include cold junction compensation error
 
T
0~  400 °C
1°C
0.1%+1
Not include cold junction compensation error
 
B
500~1800 °C
1°C
0.1%+4
Not include cold junction compensation error
RTD
Pt100
0~850°C
1°C
0.1%+1
Not contain lead resistance
RTD
Cu50
0~850°C
1°C
0.1%+1
Not contain lead resistance
Range conversion
mA
0~24.000mA
Maximum 4 minimum 1
0.025%+10
Maximum quantities: 99900
 
 
OUTPUT
Output type
Output  range
Setting step
Accuracy
Remark
DC Voltage
V
 0 ~11.00V
 0.01/0.1/1V
0.025%+1
 
DC milivolt
mV
0~110.00mV
0.1/1/10mV
0.025%+3
 
Ohm
Ω
50~400 Ω
1/10/100 Ω
0.025%+1
Not contain lead resistance
DC Current
mA
0~24.00mA
0.01/0.1/4 mA
0.025%+1
 
Thermocouple
R
0~1700 °C
1/10/100°C
0.1%+3
Not include cold junction compensation error
 
N
0~1300 °C
1/10/100°C
0.1%+3
Not include cold junction compensation error
 
S
0~1600 °C
1/10/100°C
0.1%+3
Not include cold junction compensation error
 
K
0~1370 °C
1/10/100°C
0.1%+1
Not include cold junction compensation error
 
E
0~1000 °C
1/10/100°C
0.1%+1
Not include cold junction compensation error
 
J
0~1200 °C
1/10/100°C
0.1%+1
Not include cold junction compensation error
 
T
0~  400 °C
1/10/100°C
0.1%+1
Not include cold junction compensation error
 
B
500~1800 °C
1/10/100°C
0.1%+4
Not include cold junction compensation error
RTD
Pt100
0~850°C
1/10/100°C
0.1%+1
Not contain lead resistance
RTD
Cu50
0~850°C
1/10/100°C
0.1%+2
Not contain lead resistance
24V
24V mA
24V:Can not be set Current Meassurment:0~24.000mA
           X
24V:0~-20% Current Meassurment0.025%+4
 
 

Write your review

Process calibrator: mA,V,mV,Ω,Pt100,Cu50, TC:E/K/B/S/T/R/J/N

Process calibrator: mA,V,mV,Ω,Pt100,Cu50, TC:E/K/B/S/T/R/J/N

This instrument can output/measurement the following  signals:Current/Voltage/mV; thermocouple (E,K,B,S,T,R,J,N), resistance 999.9 Ω, Pt100,Cu50, containing 14 input,15 output,analog meter site can  be read or the vast majority of quantities.V,mV,mA, Ω input and output has a  corresponding button,call directly,no combination of keys,quick and easy  operation.Small instrument,Rich functionality

Write your review